Skip to main content
HQ: Salisbury, Maryland
1-888-286-7643
Toroid Corporation
CN170812 toroidal transformer from Toroid Corporation
Max leakage <100 µAUL Recognized ANSI/AAMI ES60601-1:2005 (R)2012
01 / 02
Medical · Toroidal Magnetics

Medical IsoBox Transformers (Enclosed (white aluminum)).

Model CN170812

Enclosed hospital-grade isolation transformer (filtering & surge suppression STANDARD)

UL Recognized ANSI/AAMI ES60601-1:2005 (R)2012cUL CAN/CSA C22.2 No.60601-1 (3rd Ed.)
Rated Power
1700
VA · standard
Leakage
<100 µA
guaranteed
Dimensions
13.1 x 9.6 x 4.4
Ø × H, in

Quotes & availability:[email protected]1-888-286-7643

Technical Overview

Engineered to the stringent safety standards of UL Recognized ANSI/AAMI ES60601-1:2005 (R)2012 and international equivalents.

Enclosed hospital-grade isolation transformer (filtering & surge suppression STANDARD)

Part of the CN Series, Domestic, Single primary configuration, rated at 1700 VA.

  • 2-agency safety recognition
  • Medical-grade construction
  • 2 mounting options
  • Leakage current <100 µA
At a glance
Part Number
CN170812
Series
CN Series, Domestic
Rated Power
1700 VA
Primary
120 V
Secondary
120 V
Grade
Medical
Frequency
Domestic, 120V in / 120V out / 60 Hz
Specifications

Part-level electrical & physical data

Part NumberCN170812
SeriesCN Series, Domestic
Rated Power1700 VA
Primary Voltage120 V
Primary ConfigurationSingle
Secondary Voltage120 V
Secondary Current (Series)14.2 A
Regulation / Volt Drop<2.5%
Leakage Current<100 µA
Outlets8
Market / FrequencyDomestic, 120V in / 120V out / 60 Hz
Form FactorEnclosed (white aluminum)
GradeMedical
Dimensions (Ø × H, in)13.1 x 9.6 x 4.4
Weight (lbs)34.8
Mechanical

Top-view outline & envelope

Ø 13.1 in
Outer Ø
13.1 in
Height
9.6 in
Weight
34.8 lbs

Standard mounting: Rubber feet + brackets; cord 8 ft (up to 15). Custom brackets, potting and enclosures available on request.

Agency Recognition

Full agency recognition, ready to certify.

  • UL Recognized ANSI/AAMI ES60601-1:2005 (R)2012
  • cUL CAN/CSA C22.2 No.60601-1 (3rd Ed.)
Resources

Everything you need to specify or design-in

Talk to Engineering

Talk to a magnetics designer. Not a sales rep.

Our engineering team reviews every specification for CN170812 directly. Quotes typically turn around within 48 hours.